We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

© 2016 Wolters Kluwer Health, Inc. All rights reserved. The aim of this study was to assess the accuracy of automated blood pressure monitors on sale to the UK general public. We conducted static pressure accuracy testing on all compatible (19 out of 22 available) blood pressure monitors available for sale in pharmacies within the city of Oxford, UK, and tested two devices for accuracy in measurement of systolic and diastolic blood pressures in 21 adults. The devices showed good accuracy when measuring static pressure in laboratory bench testing, with the median error per device ranging from-2.2 to +1.2 mmHg; however, the two devices tested performed worse in vivo than in laboratory tests, with median errors as high as 6 mmHg. The monitors showed good accuracy in static pressure testing, with a lack of correlation between monitor price and accuracy. However, higher error rates seen during in-vivo testing of a subset of monitors may indicate that static testing may not be appropriate for routine accuracy assessment of these monitors.

Original publication

DOI

10.1097/MBP.0000000000000151

Type

Journal article

Journal

Blood Pressure Monitoring

Publication Date

01/01/2016

Volume

21

Pages

59 - 61