Accuracy of methods for detecting an irregular pulse and suspected atrial fibrillation: A systematic review and meta-analysis
Taggar JS., Coleman T., Lewis S., Heneghan C., Jones M.
© European Society of Cardiology 2015. Background Pulse palpation has been recommended as the first step of screening to detect atrial fibrillation. We aimed to determine and compare the accuracy of different methods for detecting pulse irregularities caused by atrial fibrillation. Methods We systematically searched MEDLINE, EMBASE, CINAHL and LILACS until 16 March 2015. Two reviewers identified eligible studies, extracted data and appraised quality using the QUADAS-2 instrument. Meta-analysis, using the bivariate hierarchical random effects method, determined average operating points for sensitivities, specificities, positive and negative likelihood ratios (PLR, NLR); we constructed summary receiver operating characteristic plots. Results Twenty-one studies investigated 39 interventions (n = 15,129 pulse assessments) for detecting atrial fibrillation. Compared to 12-lead electrocardiography (ECG) diagnosed atrial fibrillation, blood pressure monitors (BPMs; seven interventions) and non-12-lead ECGs (20 interventions) had the greatest accuracy for detecting pulse irregularities attributable to atrial fibrillation (BPM: sensitivity 0.98 (95% confidence interval (CI) 0.92-1.00), specificity 0.92 (95% CI 0.88-0.95), PLR 12.1 (95% CI 8.2-17.8) and NLR 0.02 (95% CI 0.00-0.09); non-12-lead ECG: sensitivity 0.91 (95% CI 0.86-0.94), specificity 0.95 (95% CI 0.92-0.97), PLR 20.1 (95% CI 12-33.7), NLR 0.09 (95% CI 0.06-0.14)). There were similar findings for smartphone applications (six interventions) although these studies were small in size. The sensitivity and specificity of pulse palpation (six interventions) were 0.92 (95% CI 0.85-0.96) and 0.82 (95% CI 0.76-0.88), respectively (PLR 5.2 (95% CI 3.8-7.2), NLR 0.1 (95% CI 0.05-0.18)). Conclusions BPMs and non-12-lead ECG were most accurate for detecting pulse irregularities caused by atrial fibrillation; other technologies may therefore be pragmatic alternatives to pulse palpation for the first step of atrial fibrillation screening.